ふ SK-1000WP SK-2000WP SK-5000WP SK-5001WP SK-10KWP SK-20KWP SK-30KWP 1WM PD4000461B -0-
SAFETY PRECAUTIONS All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Institute: Product Safety Signs and Labels). The meanings are as follows: A potentially hazardous situation which, if not WARNING avoided, could result in death or serious injury. A potentially hazardous situation which, if not CAUTION avoided, may result in minor or moderate injury. This manual is subject to change without notice at any time to improve the product.
PARTS DESCRIPTION Weighing Platform Spirit level ON/OFF key Turns the scale power on/off. UNITS key (only for (k)g- lb model) Changes weighing units, (k)g or lb. RE-ZERO key Display Sets display to zero or subtracts weight of a container. Four leveling feet Battery cover Batteries are not included. Battery cover lock Battery compartment AC adapter socket Battery cover lock DISPLAY STABLE indicator Indicates when the reading is stable.
PREPARATION Installing Batteries / Connecting the AC Adapter Installing Batteries CAUTION When replacing batteries, be careful not to apply too much force to the weighing platform. An excessive force could damage the scale. Unlock and remove the battery cover. (See illustration at the right) Insert six new batteries (R20P / LR20 / D size) into the battery compartment, taking extreme care to that the plus and minus polarities are observed.
WEIGHING Before Weighing Automatic Power-Off Function. If the scale is left ON and the stable indicator is displayed, the automatic power-off function turns the power off after approximately five minutes (factory setting). To disable or enable the function, press the [ON/OFF] key while pressing the [RE-ZERO] key. The scale stores this setting. To verify this, press both the [RE-ZERO] and [UNITS] keys while in the weighing mode. (See "CALIBRATION" about the [UNITS] key.
CALIBRATION When to Calibrate Calibration may be required when the scale is initially installed, if it is moved a substantial distance, or in accordance with local regulations. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur. The Position of the [UNITS] key for the Calibration The (k)g model has no [UNITS] key, but the key for the calibration is located under the overlay.
Calibration by Gravity Compensation If you have calibration weights at the capacity of the scale, you can calibrate the scale with the weights instead of compensating for the gravity acceleration. The scale is calibrated at 9.798 m/sec2 at the factory. If the gravity acceleration at your location is not the same, calibrate the scale by compensating for the gravity acceleration. (See “The Value of Gravity at Various Locations” and “World Map” and find your location value.) 1. Enter calibration mode.
SPECIFICATIONS Model SK-1000WP SK-2000WP SK-5000WP SK-5001WP SK-10KWP SK-20KWP SK-30KWP 1000 g 2000 g 5000 g 5000 g 10 kg 20 kg 30 kg Capacity 2.2 lb 4.4 lb 11 lb --22 lb 44 lb 66 lb 0.5 g 1g 2g 1g 0.005 kg 0.01 kg 0.02 kg Resolution 0.001 lb 0.002 lb 0.005 lb --0.01 lb 0.02 lb 0.05 lb Non-linearity ±1g ±2g ±4g ±4g ±0.01 kg ±0.02 kg ±0.04 kg Repeatability 0.5g 1g 2g 2g 0.005 kg 0.01 kg 0.02 kg Span drift ±0.
World Map -8-
3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148 A&D ENGINEERING, INC. 1756 Automation Parkway, San Jose, California 95131 U.S.A.