Company Ltd Instruction manual Blood Pressure Monitor TM-2655/TM-2655P

15
7-2 Description of the Functions
Table 5 Functions
No. Function SYSTOLIC
display
DIASTOLIC
display
Description
F01
f01 off
Not available.
off
No printing
1
High-speed printing
2
3-line printing
3
3-line printing
4
Graph printing
F02
f02
5
Table printing
off
No bit pattern printing
F03
Printer setting
(See “7-3
Printing
Samples” for
what each
method printing
looks like.)
TM-2655P only
f03
on
Bit pattern printing
999
The results remain in the display unless
another operation is performed.
5
Displays the results for 5 seconds.
10
Displays the results for 10 seconds.
F04 Displaying time
f04
20
Displays the results for 20 seconds.
1
Terminals 1 and 2 to a PC
2
Terminal 1 to a PC
Terminal 2 to a card reader
F05 External
connection
Ff05
3
Terminal 1 to a scale
Terminal 2 to a PC
au
Automatic pressurization
160
Pressurizes the cuff up to 160 mmHg.
180
Pressurizes the cuff up to 180 mmHg.
F06
Pressurization
Ff06
200
Pressurizes the cuff up to 200 mmHg.
120
1200 bps
240
2400 bps
480
4800 bps
F07 Channel 1 baud
rate
Ff07
960
9600 bps
120
1200 bps
240
2400 bps
480
4800 bps
F08 Channel 2 baud
rate
Ff08
960
9600 bps
F09
Ff09 H9
Not available.
off
Does not detect a human presence.
F10 Human sensor
Ff10
on
Detects a human presence.
F11 Automatic
Ff11 off
Does not output the results
automatically.