Re-calibration
Referencing the results of the self-check and using the “h” mark (the
mid-point between the highest and the lowest point among A, B, C, D).
1. Enter self-calibration mode
a. Power off the instrument and face the X-axis towards the wall.
b. Press button and button simultaneously. Then
release the power button while still pressing the vibrate button.
Release the vibrate button after all LED flashes three times. The
laser will enter self-calibration mode and continuously rotates.
2. X-axis calibration
a. Opening the cover on the remote control, press the
button, the X LED light on the laser keypad flashes
which means the laser has entered the self-calibration mode
for X-direction.
b. Press button to make the laser beam move up and
down until it coincides with the “h” mark.
3. Y-axis calibration
a. Turn the instrument by 90º in self-calibration mode and
make Y-axis facing the wall.
b. Press button, the Y LED light on the laser keypad
flashes which means the laser enters self-calibration mode for
Y-direction.
c. Press button to make the laser beam move up and
down until it coincides with the “h” mark.
4. Self-calibration confirmation
Press key after finishing re-calibration on both X and Y