User Manual

(5) Stable time
Display
Parameter
Instruction
P7 1
1
fast
2
middle
3
slow
(6) Stable range
Dispaly
Parameter
Instruction
P8 1
1
lower
2
middle
3
high
(7) Animal weighing function switch on/off
Display
Parameter
Instruction
P9 1
1
Turn off the animal function
2
Turn on the animal function
(8) RS232 output protocol selection
Display
Parameter
Instruction
P10 1
-
Reference communication data
5.
Error and correction
ERROR
Reason
Err1
The AD value is too small when calibrated or the load cell capacity is too
large
Err2
The zero point is out of the range when calibrated
Err3
The zero point is out of range upon starting or there is heavy weights
Err5
There is no load weight in calibration mode
Err7
The load cell cable connection is wrong, please check the connection method
again.
Err8
Load cell signal is not stable
Err9
The maximum range voltage is exceeded, please connect or select the
combined sensor correctly
Err10
Storage error