Adam Equipment LATITUDE Software version: 2.
Easy Reference: Model name of the scale: Serial number of the unit: Software revision number (Displayed when power is first turned on): Date of Purchase: Name of the supplier and place: © Adam Equipment Company 2018
CONTENTS P.N. 3.02.6.6. 14095, Revision D, August 2018 1.0 INTRODUCTION………………………………………………….……..…………..2 2.0 SPECIFICATIONS .............................................................................................3 3.0 SAFETY PRECAUTIONS..................................................................................3 4.0 SET UP ..............................................................................................................4 4.1 ASSEMBLING OF SCALE ..........................................
1.0 INTRODUCTION • The LBX models are supplied with a power supply adapter, to charge the internal rechargeable battery. • In use cover fitted as standard on all models. • Four models: 3kg x 0.
2.0 SPECIFICATIONS Model LBX 3 LBX 6 LBX 12 LBX 30 capacity 3Kg 6Kg 12Kg 30Kg Readability 0.5g 1g 2g 5g Repeatability 0.5g 1g 2g 5g Linearity ± 1g 2g 4g 10g Units Stabilization time Operating temperature kg / g / lb / oz / lb:oz 1 second Power supply 1 second 1 second 0ºC to 40ºC 32ºF to 104ºF 5VDC @ 800mA adapter 4V4.
Notice: • Avoid extremes of temperature, do not place in direct sunlight or near air • • • • conditioning vents. Battery should be removed if the scale is not used for a long period of time. Do not stack material on the scale when it is not in use. Avoid high humidity that might cause condensation and keep away from direct contact with water. Static may influence the weighing result, to reduce the static wipe the pan and the scale with anti-static wipes. • When the battery is lower than 3.
5.2 WEIGHING • Press [On/Off] to turn on the scale. The display will show the software revision then show the model “CAP XX”. Then it will begin its self-checking test from 0 to 9. When reading is stable, the display will show " . ", stable symbol, zero symbol, and the weighing unit used the last time the scale was used. • Press [UNITS/→] to select units from "g" / "kg" / "lb" / "oz"/”lb:oz”. Only the weighing units enabled will be displayed, See section 6.8. • Check if the reading is “ .
• Press the [Tare/↵] key once the stability indicator comes on to confirm the • • entered value. The display will show the number of pieces placed on the pan. If more items are put on the pan, the increased count value will be shown. Press [FUNC/↑] to exit and go back to the normal weighing mode. 5.3.2 PERCENTAGE WEIGHING • In the weighing mode, press [FUNC/↑] until the display shows “PER”. • Press the [Tare/↵] key to set the reference sample value, the display will show “SET”.
• the scale will show “xxxxx.x” with the ‘HI’ LED lit. Otherwise the scale will show “xxxxx.x” with the ‘OK’ LED lit. • Press the [FUNC/↑] key to exit and go back to weighing mode. 5.3.4 HOLD FUNCTION • When in weighing mode, press [FUNC/↑] until the display shows “X Hod” • Press the [UNITS/→] key to go through the holding time of 5,10,20,30, AUTO or Manual.
• To continue to other function settings, press [Tare/↵] to confirm the current setting and show the next mode. • The display will then show “BK Off”, “BK AU”, or “BK on” these are the backlight function settings. 6.3 BACKLIGHT SETTING MODE • Press the [FUNC/↑] key to change the state of backlight. • “BK Off”: the backlight is off all the time. • “BK AU”: auto backlight, backlight will come on if the display moves off of zero. • “BK on”: backlight will be on all the time.
6.6 CHECK WEIGHING FUNCTION • • • • Press [FUNC/↑] to switch on or off this function. "CHK of”:check weighing function is off. "CHK on":check weighing function is enabled. Press the [Tare/↵] key to confirm and move to the next setting. The calibration mode has been entered when “HOD on” or “HOD of” is displayed. 6.7 HOLD FUNCTION MODE • Press [FUNC/↑] to change the state of the function. • "HOD of":Hold function is off all the time and not available to the user to set as in section 5.3.4.
7.0 CALIBRATION • • • 7.1 Calibration before use is recommended for best accuracy. It is suggested that the scale be calibrated when first installed and any time the scale is moved to a different environment. Ensure weights are ready before calibrating, the weight should be of a known accuracy, for example OIML class M1 or Class 4 of ASTM E617. Do not calibrate the scale if a proper calibration weight is not available.
• If it is outside of the previous calibration value by more than 5%, it will show “FAIL L” or “FAIL H” indicating the new calibration weight was too low (FAIL L) or too high (FAIL H) in respect to the previous calibration value. 8.
WARRANTY INFORMATION Adam Equipment offers Limited Warranty (Parts and Labour) for the components failed due to defects in materials or workmanship. Warranty starts from the date of delivery. During the warranty period, should any repairs be necessary, the purchaser must inform its supplier or Adam Equipment Company. The company or its authorized Technician reserves the right to repair or replace the components at any of its workshops depending on the severity of the problems.
WEEE 2012/19/EU This device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. Disposal of batteries (if fitted) must conform to local laws and restrictions. Cet appareil ne peut être éliminé avec les déchets ménagers. L’élimination de la batterie doit être effectuée conformément aux lois et restrictions locales. Dieses Gerät nicht mit dem Hausmüll entsorgt.
ADAM EQUIPMENT is an ISO 9001:2008 certified global company with more than 40 years’ experience in the production and sale of electronic weighing equipment. Adam products are predominantly designed for the Laboratory, Educational, Health and Fitness, Retail and Industrial Segments.