Product Manual

38 | P a g e © Adam Equipment Company 2010
13.0 ERROR
CODES
ERROR
CODES
DESCRIPTION SUGGESTIONS
- -oL - -
Over-range Remove weight from the scale.
If the problem persists contact your dealer or Adam
Equipment for assistance.
Err 1
Time Setting Error
Enter time using correct format and reasonable values.
Format: hh:mm:ss
Err 2
Date Setting Error Enter date using correct format and reasonable values.
Format: yy:mm:dd
Err 4
Zero Setting Error The scale was outside the normal zero setting range
either when it was turned on or when the [Zero] key
was pressed.
Remove weight from the scale and try re-zeroing again.
Use the [Z/T] key to set the display to zero value.
If the problem persists contact your dealer or Adam
Equipment for assistance.
Err 6
A/D out of range The values from the A/D converter are outside the
normal range.
Remove the weight from the scale if overloaded.
Make sure the pan is attached.
Indicates the load cell or the electronics may be faulty.
If the problem persists contact your dealer or Adam
Equipment for assistance.
Err 9
Check weigh
limits error
Shown if the low limit is set higher than the current
high limit. Reset High limit or different low limit.
FAIL
Calibration error. Improper calibration (should be within + 10% of the
factory calibration). The old calibration data will be
retained until the calibration process is complete.
If the problem persists contact your dealer or Adam
Equipment for assistance.