The auto power off function helps conserve power by automatically turning the
balance off when it is not in use. The Auto power off time can be set by the user in
minutes, or to not be in operation (OFF).
1) Switch on the balance then press and hold the [Unit/Clr] key during the self-
checking test of the display.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Air/Liquid] key until F5 OFF is displayed.
4) Press the [Tare] key to view the current setting.
5) Press [Air/Liquid] to change the setting to (Off, 5, 10, 20 or 30 minutes).
6) Press [Tare] to confirm and move to the next setting.
7) When F5 oFF is displayed press the [Tare] key to enter the parameter, the
[Unit/Clr] key to return to weighing, or press [Air/Liquid] to go to the next
function.
7.6 INTERNAL
OR
EXTERNAL
CALIBRATION,
F6
F6F6
F6
I
II
ICE
CECE
CE
You can select to use the internal mass or an external mass to calibrate the balance.
1) Switch on the balance then press and hold the [Unit/Clr] key during the self-
checking test of the display.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Air/Liquid] key until F6 ICE is displayed.
4) Press the [Tare] key to view the current setting.
5) To change the setting press the [Air/Liquid] key to select either Int
CAL
(internal calibration) or Etl
CAL
(external calibration).
6) Press [Tare] to confirm and move to the next setting.
7) When F7 CA is displayed press the [Tare] key to enter the parameter, the
[Unit/Clr] key to return to weighing, or press [Air/Liquid] to go to the next
function.