User Guide
18 Clamp-on Ground Resistance Tester Model 6416
Menu # Indication Function
4 AL. Ω
Setting the Impedance Alarm Threshold (Ω):
■ Enter the menu by ►. W blinks.
■ Press ▲ or ▼ to select the state of the alarm:
-
: Disabled.
-
:
Enabled for a measurement exceeding the
threshold.
-
:
Enabled for a measurement below the
threshold.
■ Validate by ►.
Setting the Alarm Value:
■ Press ▲ or ▼ to select the impedance alarm
threshold (Figure 2-1, item #11).
■ Validate and return to the previous menu by ►.
5 AL. V
Setting the Voltage Alarm Threshold (V):
■ Enter the menu by ►. blinks.
■ Press ▲ or ▼ to select the state of the alarm
(Figure 2-1, item #10):
-
: Disabled.
-
:
Enabled for a measurement exceeding the
threshold.
■ Validate by ►.
Setting the Alarm Value.
■ Press ▲ or ▼ to select the alarm threshold value
(Figure 2-1, item #11).
■ Validate and return to the previous menu by ►.
6 AL. A
Setting the Current Alarm Threshold (I):
■ Enter the menu by ►. blinks.
■ Press ▲ or ▼ to select the state of the alarm
(Figure 2-1, item #10):
-
: Disabled.
-
:
Enabled for a measurement exceeding the
threshold.
■ Validate by ►.
Setting the Alarm Value.
■ Press ▲ or ▼ to select the current alarm threshold
(Figure 2-1, item #11).
■ Validate and return to the previous menu by ►.