User Guide

26 Clamp-on Ground Resistance Tester Model 6416
5.8.4 No Alarm Detection
If no alarm is activated, the alarm icons are not displayed.
When no alarm is triggered, the
alarm threshold is displayed,
along with the direction
of triggering (
, ) of the
impedance, voltage, or current
alarm.

5.8.5 Priority of the Alarms
If several alarms are triggered simultaneously, a priority rule determines the
display and the corresponding sound:
The voltage alarm has priority because it concerns the user’s safety.
The current alarm is second in priority.
The impedance alarm is displayed when no other alarm is triggered.