User manual

REMOTE INTERFACE
99
This means that the magnitude of the DC offset is
determined by the output amplitude.
If the specified DC offset is out of range, the
maximum/minimum offset will be set instead. A
-222 error will be generated from the remote
terminal.
SOURce[1]:APPLy:SINusoid
Set
Outputs a sine wave when the command has
executed. Frequency, amplitude and offset can also
be set.
SOURce[1]:APPLy:SINusoid [<frequency>
[,<amplitude> [,<offset>] ]]
<frequency>
0.1Hz~25MHz*
<amplitude>
1mV~10Vpp (50)
<offset>
-5V ~ +5V (50)
*AFG-2005/2105 limited to 5MHz, AFG-2012/2112
limited to 12MHz.
SOURce1:APPL:SIN MAX, 3.0, -2.5
Outputs a 3Vpp sine wave at 25MHz (max
frequency) with a -2.5V offset.
SOURce[1]:APPLy:SQUare
Set
Outputs a square wave when the command has
executed. Frequency, amplitude and offset can also
be set. The duty cycle is fixed to 50%.
SOURce[1]:APPLy:SQUare [<frequency> [,<amplitude>
[,<offset>] ]]
<frequency>
0.1Hz ~ 25MHz*
<
amplitude
>
1mV~10V (50)