User manual

AFG-2000 Series User Manual
126
SOURce[1]:SWEep:STATe
Set
Query
Description
Sets or disables Sweep mode. By default sweep is
disabled. Sweep must be enabled before setting
other parameters.
Note
Any modes will be disabled if sweep mode is
enabled.
Syntax
SOURce[1]:SWEep:STATe {OFF|ON}
Example
SOUR1:SWE:STAT ON
Enables sweep mode.
Query Syntax
SOURce[1]:SWEep:STATe?
Return Parameter
0
Disabled (OFF)
1
Enabled (ON)
Query Example
SOUR1:SWE:STAT?
>1
Sweep mode is currently enabled.
SOURce[1]:FREQuency:STARt
Set
Query
Description
Sets the start frequency of the sweep. 100Hz is the
default start frequency.
Note
To sweep up set the stop frequency higher than the
start frequency. Set the stop frequency lower than
the start frequency to sweep down.
Syntax
SOURce[1]:FREQuency:STARt
{<frequency>|MINimum|MAXimum}
Parameter
<frequency>
0.1Hz ~ 25MHz*
0.1Hz ~ 1MHz (Ramp)
*AFG-2112 limited to 12MHz, AFG-2105 limited to
5MHz.