Save and Recall Commands
Up to 10 different instrument states can be stored to non-volatile
memory (# 0~9) and up to 10 different ARB waveforms can be saved
to memory locations 10~19.
*SAV
Set
Saves the current instrument state to a specified
save location or an ARB waveform to the specified
location. When a state is saved, all the current
instrument settings, functions, modulation
parameters and waveforms are also saved.
Memory locations 0~9, save the instrument state
only, whilst memory locations 10~19 save ARB
data.
The *RST command will not delete saved
instrument states from memory.
*SAV {NR1}
0~9
Save state
10~19
Save ARB data
*SAV 0
Save the instrument state to memory location 0.
*RCL
Set
Recall previously saved instrument states from
memory locations 0~9 or recall the previously
saved ARB waveforms from memory locations
10~19.
*RCL {NR1}
0~9
Recall state
10~19
Recall ARB data