User Manual

Table Of Contents
47
2921B EN 20031206
Establishing the daily operating reference density levels
This procedure enables you to establish the base line values for:
Base+Fog density,
Mid density,
DD (Density Difference),
MaxD (Maximum Density).
To establish the daily operating density levels, proceed as follows:
1 Press the Key-operator key to enter the key-operator mode.
2 Press the Down key seven times, followed by the Confirm key to select
‘Quality control’.
A confirmation screen appears:
3 Press the Confirm key (YES) to start the QC procedure or the Escape key
(NO) to quit.
The Drystar 5300 will automatically print the QC test image.
By default, the internal densitometer of the Drystar 5300 will be used. The
internal densitometer of the Drystar 5300 is calibrated at installation.
Authorized service personnel should recalibrate the internal
densitometer annually or after major service or repair.
Do not change the ‘default’ internal densitometer during the entire QC
procedure!
1 Show settings
2 Change settings
3 Print image
4 Save configuration
5 Restore config.
6 Calibration
7 Service actions
8 Quality control
9 Installation
KO
Start QC
Procedure
YES (v) NO (x)
QC
QC procedure
in progress
Please wait
62%
QC