100 Chapter 7
COM Object Reference
SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. SAVE
SCPI.SENSe(Ch).CORRection.MULTiport.COLLect.
SAVE
Object type Property
Syntax SCPI.SENSe(Ch).CORRection.MULTiport.COLLect.SAVE
Description For the active trace of channels 1 to 4 (Ch), calculates the calibration coefficients
depending on the selected calibration type. Calibration coefficient is calculated for
all port pairs that have all necessary measured calibration data for calculation. If
none of the port pair has all necessary data for calculation, an error occurs and the
command is ignored. The command clears all measured calibration data of the
channel. (No query)
NOTE The same calibration method must be used for all port pairs to be used in a channel.
Therefore, if calibration is executed after changing the calibration method by
SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. METHod on page 98, all
the calibration coefficient for the different calibration method will be discarded.
Variable For information on the variable (Ch), see Table 7-2, “Variable (Ch),” on page 91.
Examples
SCPI.SENSe(1).CORRection.MULTiport.COLLect.SAVE
Related objects SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.LOAD on
page 93
SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.OPEN on
page 94
SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.SHORt on
page 95
SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.THRU on
page 96
SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. METHod on page 98
Equivalent key
[CAL] - Test Set Cal - Calibration - Done