user manual

102 Chapter 7
COM Object Reference
SCPI.SENSe(Ch).CORRection.MULTiport.SELF. STATe
SCPI.SENSe(Ch).CORRection.MULTiport.SELF.
STATe
Object type Property
Syntax SCPI.SENSe(Ch).CORRection.MULTiport.SELF.STATe = Status
Status = SCPI.SENSe(Ch).CORRection.MULTiport.SELF.STATe
Description For channel 1 to channel 4, specifies whether to execute the self calibration of the
test set automatically at user defined interval. To specify the interval, use the
SCPI.SENSe(Ch).CORRection.MULTiport.SELF. TIMer on page 103
command. Note that, if SCPI.SENSe(Ch).MULTiport.STATe on page 107 is off,
self calibration is not executed automatically even if it is turned on with this
command.
Variable
For information on the variable (Ch), see Table 7-2, “Variable (Ch),” on page 91.
Examples
Dim Status as Boolean
SCPI.SENSe(1).CORRection.MULTiport.SELF.STATe = True
Status = SCPI.SENSe(1).CORRection.MULTiport.SELF.STATe
Related objects SCPI.SENSe(Ch).CORRection.MULTiport.SELF. ONCE on page 101
SCPI.SENSe(Ch).CORRection.MULTiport.SELF. TIMer on page 103
SCPI.SENSe(Ch).MULTiport.STATe on page 107
Equivalent key
[CAL] - Test Set Cal - Self Cal
Status
Description Auto Self Cal state
Data type Boolean
Range True (-1) or False (0)
Preset value True