Chapter 1 13
Introduction and Installation
Installation
NOTE There are special semi-rigid cables that should be used to connect the analyzer’s test
ports to the test set’s REFLECTION and TRANSMISSION ports when used in a
rack (or a bench configuration where the analyzers bottom feet have been
removed). These cables were shipped with your analyzer if you ordered Option
1CM. If you order the rack mount kit separately, you will need to order these special
cables separately also. See Table 1-1 for part numbers.
To install your system in an 85043D rack, follow the instructions in the rack
manual.
CAUTION When installing your system in other racks, improper installation may cause shock
hazards, overheating, dust contamination, and inferior system performance. For
support details and information about installation and warranty, call Agilent
Technologies. See page 120 for contact information.
CAUTION When installing the system in a cabinet, the convection into and out of the system
must not be restricted. The ambient temperature (outside the cabinet) must be less
than the maximum operating temperature of the system by 4 °C for every 100 watts
dissipated in the cabinet. If the total power dissipated in the cabinet is greater than
800 watts, then forced convection must be used.