26 Chapter 2
Making Measurements
Measurement Procedures
Step 4. Executing calibration
After the measurement of the calibration standards for necessary ports is finished,
calculate and save the calibration coefficients.
[Cal] - Test Set Cal - Calibration - Done
NOTE The calibration data of the 87075C multi-port test set has higher priority over the
analyzers normal calibration data. Therefore, if Test Set Cal is turned on with
normal calibration data existed, the multi-port test set calibration data is used, and
the normal calibration data is ignored.