user manual

Chapter 6 71
SCPI Command Reference
:SENS{1-4}:CORR:MULT:COLL:ISOL
:SENS{1-4}:CORR:MULT:COLL:ISOL
Syntax :SENSe{[1]|2|3|4}:CORRection:MULTiport:COLLect[:ACQuire]:ISOLation <value 1>,<value 2>
Description For channel 1 to channel 4, measures isolation calibration data for the specified port
pair.
NOTE If 2 ports you specify are not a port pair, an error occurs. Also, if
:SENS{1-4}:MULT on page 81 is off, an error occur.
Parameters
Query response No query
Example of use
10 OUTPUT 717;":SENS1:CORR:MULT:COLL:ISOL 1,2"
Related commands :SENS{1-4}:MULT on page 81
:SENS{1-4}:CORR:MULT:COLL:OPEN on page 74
:SENS{1-4}:CORR:MULT:COLL:SHOR on page 76
:SENS{1-4}:CORR:MULT:COLL:LOAD on page 72
:SENS{1-4}:CORR:MULT:COLL:THRU on page 77
Equivalent key
[Cal] - Test Set Cal - Calibration - Port x-y - Port x-y Isol (Optional)
Description
value 1 Specifies a port of the test set port pair for which isolation
calibration data is measured.
value 2 Specifies the other port of the test set port pair for which
isolation calibration data is measured.