SCPI Command Reference
:SENS{1-4}:CORR:MULT:COLL:SAVE
:SENS{1-4}:CORR:MULT:COLL:SAVE
Syntax :SENSe{[1]|2|3|4}:CORRection:MULTiport:COLLect:SAVE
Description For channel 1 to channel 4, calculates the calibration coefficients depending on the
selected calibration type. Calibration coefficient is calculated for all port pairs that
have all necessary measured calibration data for calculation. If none of the port pair
has all necessary data for calculation, an error occurs and the command is ignored.
The command clears all measured calibration data of the channel. (No query)
NOTE The same calibration method must be used for all port pairs to be used in a channel.
Therefore, if calibration is executed after changing the calibration method by
:SENS{1-4}:CORR:MULT:COLL:METH on page 73, all the calibration
coefficient for the different calibration method will be discarded.
Query response No query
Example of use
10 OUTPUT 717;":SENS1:CORR:MULT:COLL:SAVE"
Related commands :SENS{1-4}:CORR:MULT:COLL:METH on page 73
Equivalent key
[Cal] - Test Set Cal - Calibration - Done