76 Chapter 6
SCPI Command Reference
:SENS{1-4}:CORR:MULT:COLL:SHOR
:SENS{1-4}:CORR:MULT:COLL:SHOR
Syntax :SENSe{[1]|2|3|4}:CORRection:MULTiport:COLLect[:ACQuire]:SHORt <value>
Description For channel 1 to channel 4, measures the calibration data of the short standard of the
specified port.
NOTE If :SENS{1-4}:MULT on page 81 is off, an error occur.
Parameters
Query response No query
Example of use
10 OUTPUT 717;":SENS1:CORR:MULT:COLL:SHOR 1"
Related commands :SENS{1-4}:MULT on page 81
:SENS{1-4}:CORR:MULT:COLL:OPEN on page 74
:SENS{1-4}:CORR:MULT:COLL:LOAD on page 72
:SENS{1-4}:CORR:MULT:COLL:THRU on page 77
:SENS{1-4}:CORR:MULT:COLL:ISOL on page 71
Equivalent key
[Cal] - Test Set Cal - Calibration - Port x-y - Port x Short (or Port y Short)
Description
value Specifies a port of the test set for which the calibration data
of the short standard is measured.