78 Chapter 6
SCPI Command Reference
:SENS{1-4}:CORR:MULT:SELF
:SENS{1-4}:CORR:MULT:SELF
Syntax :SENSe{[1]|2|3|4}:CORRection:MULTiport:SELF[:STATe] {ON|OFF|1|0}
:SENSe{[1]|2|3|4}:CORRection:MULTiport:SELF[:STATe]?
Description For channel 1 to channel 4, specifies whether to execute the self calibration of the
test set automatically at user defined interval. To specify the interval, use the
:SENS{1-4}:CORR:MULT:SELF:TIM on page 80 command. Note that, if
:SENS{1-4}:MULT on page 81 is off, self calibration is not executed automatically
even if it is turned on with this command.
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:MULT:SELF ON"
20 OUTPUT 717;":SENS1:CORR:MULT:SELF?"
30 ENTER 717;A
Related commands :SENS{1-4}:MULT on page 81
:SENS{1-4}:CORR:MULT:SELF:TIM on page 80
Equivalent key
[Cal] - Test Set Cal - Self Cal
Description
ON or 1 Executes self calibration for the test set at user defined
interval.
OFF or 0 (preset value) Does not execute self calibration for the test set at user
defined intervals.