Product Manual

MN164-1 Rev -
08/09
20
Classification
Type of protection against electric shock:
Class II Protection from electric shock is achieved by double insulation.
Protective earthing or reliance upon installation conditions are not required.
Degree of protection against electric shock:
Type B Equipment providing a particular degree of protection against electric shock particularly regarding:
1) allowable leakage current;
2) reliability of protective earth connection (if present).
Not intended for direct cardiac application.
Degree of protection against harmful ingress of water:
Drip-proof equipment – IPX1.
Equipment provided with an enclosure preventing of such an amount of falling liquid as might interfere with the
satisfactory and safe operation of the equipment.
Method of cleaning and infection control allowed:
Please refer to Maintenance section in the NewLife Intensity Service Manual.
Degree of safety of application in the presence of flammable anesthetic gases:
Equipment not suited for such application.
Mode of operation:
Continuous duty.