Product Manual

MN164-1 Rev -
08/09
23
Setting the Pediatric Flowmeter
When using the pediatric flowmeter, the unit will not reach proper concentration at the pediatric setting (less than 2 lpm) until
you bleed off a portion of the oxygen by opening the primary flowmeter (on the left side of the unit). Follow the procedure
below when using the pediatric flowmeter.
1. Follow the start-up instructions 1-7 as outlined on page 8 & 9.
2. Set the pediatric flowmeter to the prescribed flow.
3. Set the primary flowmeter to 2 lpm to bleed off excess product, and allow the unit to achieve maximum concentration.
The NewLife Intensity Oxygen Concentrator must be operated for
at least five minutes at 2 lpm before using the unit.
The NewLife Intensity is appropriate for usage by two patients,
provided the combined flow is a minimum of 2 lpm and does not
exceed the maximum capacity of the concentrator.
Figure 13