Replacing Halogen lamp:
Remove the Ophthalmoscope head by turning counterclockwise. Grab end of lamp from inside base
of Ophthalmoscope head and remove. Replace with new lamp (Item #5210-4), taking care not to
touch the glass, and align lamp metal guides on side of lamp with grooves in lamp socket. (Fig. 10).
Bent Arm Illuminator Head: (Item #5215-TH)
This instrument head is designed for examination of the oral cavity in conjunc-
tion with the tongue depressor blade, or laryngeal mirror attachments. Supplied
attachments include a reusable tongue blade (Item #5215-TD), adapter for dis-
posable tongue blade (Item #5215-TO), 3mm laryngeal mirror (Item #5215-
M3), and 4mm laryngeal mirror (Item #5215-M4). These attachments are supplied with complete set
(Item #5215) only or can be purchased separately. (Fig. 11).
Attaching any of the supplied Attachments
Align slot on attachment with pin on top surface of bent arm illuminator and push until secure.
(Fig. 11).
Replacing lamp (Item #5211-4) Remove by turning counterclockwise. Reverse procedure to replace.
Nasal Specula Attachment: (provided on Item #5215)
This instrument head attachment (Item #5215-NS) is designed for examination of the nasal cavity. It
is used in conjunction with the Otoscope instrument head.
Attaching Nasal Specula
Attach Otoscope head to the battery handle. Remove ear speculum or adapter.
Attach Nasal Speculum by pushing the flanged end of the speculum onto the in-
strument head aligning the slot in the speculum flange with the rivet at the top of
the instrument head and twist clockwise to engage. Reverse procedure to re-
move. (Fig. 12).
Cleaning, Disinfection, and Drying
Note: Prior to performing any cleaning or disinfection of your device, be sure to remove the batteries
from the handles. Do not re-insert the batteries without first consulting the ‘drying’ section below.
The cleaning and disinfection procedures outlined below are the only methods validated for this de-
vice. ADC does not recommend the use of ultrasonic cleaning equipment, automated washers/disin-
fectors, autoclave devices, or other cleaning methods not specified below. These alternate
cleaning/disinfecting methods have not been validated with ADC’s devices and may cause damage or
corrosion to your medical device.
All devices and accessories (excluding disposable specula) should be thoroughly cleaned and disin-
fected prior to their first use. Devices and accessories being sent for repair or servicing must also be
disinfected prior to shipping.
OUTLINE OF PROPER CLEANING/DISINFECTION:
In order to achieve proper cleaning and high level disinfection of your device, you must perform the