22. CHECKS AND ADJUSTMENTS
With this option you can measure collimation error in your instrument so that the OS can correct
subsequent single face observations. To measure the error, make angular observations using
both faces.
PROCEDURE
1.
Select "Inst. cons." in <Configuration>.
2. Select "Collimation".
3. Sight the reference point in Face 1 and press
[OK]. Telescope rotates and vertical circle is
indexed.
4. Sight the reference point in Face 2 and press
[OK].
22.3
Collimation