This instrument is designed to perform accurate level
measurements at frequencies greater than 550 MHz. In order to
achieve the desired accuracy at higher frequencies, a precision F-
connector (2110-06-1029) is installed, and a spare is provided.
Replace this connector only with a JDSU approved precision F-
connector.
The life of the F-connector will be extended by keeping a test lead
attached to the instrument at all times. Avoid using “quick F”
connectors or “push-ons” as these are likely to damage the
connector. Measurement accuracy at these frequencies is
enhanced by using a probe prescribed by amplifier manufacturer .
When a probe is not available, a 6 dB pad with good response
characteristics throughout the measurement frequency range is
recommended at the test point.
For more information about testing at higher frequencies, please
refer to JDSU’s application note, “Considerations in Testing 750/862
MHz Amplifiers”,
call 1-800-622-5515, or
e-mail to: catvsupport@JDSU.com.