User`s manual

<7. Operation>
7-8
IM 01C25B01-01E
7.6.4 TagNumberConguration
Tag Number is edited by turning the external
adjustment screw. Up to 8 alphanumeric characters
for HART or 16 alphanumeric characters for BRAIN
can be set.
F0712.ai
screw
push
Change the first Character
Go to the next Character
Set all other
characters in the
same way
Blinking Blinking
Save ?/
Cancel ?
7.6.5 PressureUnitConguration
Pressure unit for the below table can be changed
as below. By turning the external adjustment screw,
user can scroll between the various available
pressure units.
F0713.ai
screw
push
Select by the push button
Blinking Blinking
Save ?/
Cancel ?
[Available pressure units]
kPa bar inH
2
O@4degC(39.2degF)
Torr mbar inH
2
O@20degC(68degF)
atm g/cm
2
inHg
MPa kg/cm
2
ftH
2
O@4degC(39.2degF)
hPa Pa ftH
2
O@20degC(68degF)
mmHg
mmH
2
O@4degC(39.2degF)
psi
mmH
2
O@20degC(68degF)
In addition to the above units, mmAq@4degC,
mmAq@20degC, mmWG@4degC and mmWG@20degC
are available for BRAIN communication type.
7.6.6 PressureLRV/URVConguration
Pressure LRV and URV can be set. The number
for each digit is changed by turning the external
adjustment screw and set by pressing the push
button. Please refer to 7.6.7 Damping Time
ConstantCongurationforhowtochangethe
numerical value.
When the setting is out of the limit, an alarm will be
generated.