User`s manual
<8. Maintenance>
8-5
IM 01C25B01-01E
■RemovingtheCapsuleAssembly
IMPORTANT
Exercise care as follows when cleaning the
capsule assembly.
• Handlethecapsuleassemblywithcare,and
be especially careful not to damage or distort
the diaphragms that contact the process
uid.
• Donotuseachlorinatedoracidicsolutionfor
cleaning.
• Rinsethoroughlywithcleanwaterafter
cleaning.
1) Remove the CPU assembly as shown in
subsection 8.4.2.
2) Remove the two setscrews that connect the
transmitter section and pressure-detector
section.
3) Remove the hexagon-head screw and the
stopper.
4) Separate the transmitter section and pressure-
detector section.
5) Removethenutsfromthefourangebolts.
6) While supporting the capsule assembly with
onehand,removethecoverange.
7) Remove the capsule assembly.
8) Clean the capsule assembly or replace with a
new one.
■ReassemblingtheCapsuleAssembly
1) Insertthecapsuleassemblybetweentheange
bolts, paying close attention to the relative
positions of the H (high pressure side) and
L (low pressure side) marks on the capsule
assembly.
Replace the two capsule gaskets with new
gaskets.
2) Installthecoverangeonthehighpressure
side,anduseatorquewrenchtotightenthe
fournutsuniformlytoatorqueshownbelow.
Model
EJ110, EJ310, EJ430
EJ120
Wetted parts material code
S H,M,T,A,D,B
Torque(N·m)
{kgf·m}
17*
{1.7}
40
{4.1}
40
{4.1}
*: 40 for measurement span code F.
Model
EJ130 EJ440
Bolts & nuts material code
G, C J G, C J
Torque(N·m)
{kgf·m}
90
{9.2}
110
{11.2}
70
{7.1}
100
{10.2}
3) After the pressure-detector section has been
reassembled, a leak test must be performed to
verify that there are no pressure leaks.
4) Reattach the transmitter section to the
pressure-detector section.
Reattach the stopper with the hexagon-head
screw.
5) Tighten the two setscrews. (Tighten the screws
toatorqueof1.5N·m)
6) Install the CPU assembly according to
subsection 8.4.2.
7) After completing reassembly, adjust the zero
point and recheck the parameters.
F0804.ai
Transmitter section
Pressure-detector section
Cover flange
Capsule gasket
Flange bolt
Setscrew
Nut
Figure 8.4 Removing and Mounting the Pressure-
detector Section
IMPORTANT
If you remove the vent/drain plugs when
disassembling EJ130 and EJ440
transmitters, install the plugs again before
reassemblingthecoverangeswiththecapsule
assmbly.