User`s manual

<8. Maintenance>
8-6
IM 01C25B01-01E
8.4.4 Replacing the Process Connector
Gaskets
This subsection describes process connector
gasketreplacement.(Seegure8.5.)
(a) Loosen the two bolts, and remove the process
connectors.
(b) Replace the process connector gaskets.
(c) Remount the process connectors. Tighten the
boltssecurelyanduniformlytoatorqueshown
below, and verify that there are no pressure
leaks.
Model
EJ110
EJ120
EJ130
EJ310
EJ430
EJ440
C capsule D capsule
Torque(N·m)
{kgf·m}
39 to 49
{4to5}
49 to 59
{5to6}
F0805.ai
Bolt
Process connector
Process connector gasket
Figure 8.5 Removing and Mounting the Process
Connector
8.5 Troubleshooting
If any abnormality appears in the measured values,
usethetroubleshootingowchartbelowtoisolate
and remedy the problem. Since some problems
havecomplexcauses,theseowchartsmay
notidentifyall.Ifyouhavedifcultyisolatingor
correcting a problem, contact Yokogawa service
personnel.
8.5.1 BasicTroubleshooting
First determine whether the process variable
is actually abnormal or a problem exists in the
measurement system.
If the problem is in the measurement system,
isolate the problem and decide what corrective
action to take.
Thistransmitterisequippedwithaself-diagnostic
function which will be useful in troubleshooting,
andthetransmitterequippedwithanintegral
indicator will show an alarm code as a result of self-
diagnosis.
See subsection 8.5.3 for the list of alarms.
See also each communication manual.
Abnormalities appear in measurement.
: Areas where self-diagnostic offers support
Is process variable
itself abnormal?
Inspect the
process system.
Isolate problem in
measurement system.
Does problem exist in
receiving instrument?
Check/correct
environmental conditions.
Inspect receiver.
Check transmitter.
Check/correct operating
conditions.
Measurement system problem
F0806.ai
YES
NO
NO
YES
Environmental conditions
Operating conditions
Transmitter itself
Figure 8.6 Basic Flow and Self-Diagnostics