User`s manual
<8. Maintenance>
8-6
IM 01C25B01-01E
8.4.4 Replacing the Process Connector
Gaskets
This subsection describes process connector
gasketreplacement.(Seegure8.5.)
(a) Loosen the two bolts, and remove the process
connectors.
(b) Replace the process connector gaskets.
(c) Remount the process connectors. Tighten the
boltssecurelyanduniformlytoatorqueshown
below, and verify that there are no pressure
leaks.
Model
EJ110
EJ120
EJ130
EJ310
EJ430
EJ440
C capsule D capsule
Torque(N·m)
{kgf·m}
39 to 49
{4to5}
49 to 59
{5to6}
F0805.ai
Bolt
Process connector
Process connector gasket
Figure 8.5 Removing and Mounting the Process
Connector
8.5 Troubleshooting
If any abnormality appears in the measured values,
usethetroubleshootingowchartbelowtoisolate
and remedy the problem. Since some problems
havecomplexcauses,theseowchartsmay
notidentifyall.Ifyouhavedifcultyisolatingor
correcting a problem, contact Yokogawa service
personnel.
8.5.1 BasicTroubleshooting
First determine whether the process variable
is actually abnormal or a problem exists in the
measurement system.
If the problem is in the measurement system,
isolate the problem and decide what corrective
action to take.
Thistransmitterisequippedwithaself-diagnostic
function which will be useful in troubleshooting,
andthetransmitterequippedwithanintegral
indicator will show an alarm code as a result of self-
diagnosis.
See subsection 8.5.3 for the list of alarms.
See also each communication manual.
Abnormalities appear in measurement.
: Areas where self-diagnostic offers support
Is process variable
itself abnormal?
Inspect the
process system.
Isolate problem in
measurement system.
Does problem exist in
receiving instrument?
Check/correct
environmental conditions.
Inspect receiver.
Check transmitter.
Check/correct operating
conditions.
Measurement system problem
F0806.ai
YES
NO
NO
YES
Environmental conditions
Operating conditions
Transmitter itself
Figure 8.6 Basic Flow and Self-Diagnostics