<9.GeneralSpecications>
9-3
IM 01C25B01-01E
Local Parameter Setting (Output signal code D,
J and Q)
Parametercongurationbytheexternalzero
adjustment screw and push button (Integral
indicatorcodeE)offerseasyandquicksetup
for parameters of Tag number, Unit, LRV, URV,
Damping,Outputmode(linear/squareroot),
Display out 1, and Re-range by applying actual
pressure (LRV/URV).
Burst Pressure Limits
[EJ110, EJ310, EJ430]
69 MPa (10000 psi)
[EJ130, EJ440]
132 MPa (19100 psi)
Normal Operating Condition
(Selected features may affect limits.)
AmbientTemperatureLimits
–40 to 85°C (–40 to 185°F)
–30 to 80°C (–22 to 176°F) with LCD display
–25 to 80°C (–40 to 176°F) for EJ120
Process Temperature Limits
–40 to 12C (–40 to 248°F)
–25 to 80°C (–40 to 176°F) for EJ120
–40 to 10C (–40 to 212°F) for EJ310
L capsule
AmbientHumidityLimits
0 to 100% RH
WorkingPressureLimits(Siliconeoil)
EJ120 -50 to 50 kPa
Maximum Pressure Limits
EJX110A
Capsule Pressure
F, L 16 MPa (2300 psi)
M, H, V 25 MPa (3600 psi)*
* 16 MPa for wetted parts material code H, M, T, A, D,
and B.
EJA110E
Capsule Pressure
F, L, M, H, V 16 MPa (2300 psi)*
*25MPa(3600psi)when/HGisspecied.
EJ130
Capsule Pressure
M, H 32 MPa (4500 psi)
EJ310
Capsule Pressure
L 10 kPa abs (2.95 inHg abs)
M 130 kPa abs (38 inHg abs)
A 3.5 MPa abs (500 psia)
B 16 MPa abs (2300 psia)
EJ430
Capsule Pressure
H 500 kPa (2000 inH
2O)
A 3.5 MPa (500 psi)
B 16 MPa (2300 psi)
EJ440
Capsule Pressure
C 32 MPa (4500 psi)
D 50 MPa (7200 psi)
Minimum Pressure Limit
See graph below
EJ110, EJ130, EJ430, EJ440
Atmospheric
pressure
-40
(-40)
0
(32)
40
(104)
80
(176)
120
(248)
1(0.14)
2.7(0.38)
10(1.4)
(psi abs)
100(14.5)
Process temperature °C (°F)
Working
pressure
kPa abs
Applicable range
F0901.ai