DN:ZRY721-SMSY01 V1.0 Operation Guide MODEL BC 57 (KD-721) Fully Automatic Wrist Cuff Blood Pressure Monitor (ELECTRONIC SPHYGMOMANOMETER) OPERATION GUIDE __________________________________________________________________________ INDEX IMPORTANT INFORMATION ......................................................................... 2 CONTENTS AND DISPLAY INDICATORS ..................................................... 2 INTENDED USE .........................................................................
DN:ZRY721-SMSY01 V1.0 Operation Guide IMPORTANT INFORMATION NORMAL BLOOD PRESSURE FLUCTUATION All physical activity, excitement, stress, eating, drinking, smoking, body posture and many other activities or factors (including taking a blood pressure measurement) will influence blood pressure value. Because of this, it is mostly unusual to obtain identical multiple blood pressure readings. Blood pressure fluctuates continually ----- day and night.
DN:ZRY721-SMSY01 V1.0 Operation Guide INTENDED USE Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. CONTRAINDICATION It is inappropriate for people with serious arrhythmia to use this Electronic Sphygmomanometer.
DN:ZRY721-SMSY01 V1.0 Operation Guide SPECIFICATIONS 1. Product name: Blood Pressure Monitor 2. Model: BC 57 (KD-721) 3. Classification: Internally powered, Type BF applied part,IPX0,No AP or APG,Continuous operation 4. Machine size: Approx. 78mm×60mm×28mm 5. Cuff circumference: 14cm ~ 19.5cm(5 1/2″~ 7 11/16″) 6. Weight: Approx. 67g (2 3/8oz.) (exclude batteries) 7. Measuring method: Oscillometric method, automatic inflation and measurement 8. Memory volume: 2×60 times with time and date stamp 9.
DN:ZRY721-SMSY01 V1.0 Operation Guide 1) The application of the cuff over a wound or inflammation diseases; 2) The application of the cuff on any limb where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present; 3) The application of the cuff on the wrist on the side of a mastectomy; 4) Simultaneously used with other monitoring medical equipments on the same limb; 5) Need to check the blood circulation of the user. 8.
DN:ZRY721-SMSY01 V1.0 Operation Guide 16. This device complies with Part 15 of the FCC Rules. Its operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. 17.Attention that changes or modification not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. 18.
DN:ZRY721-SMSY01 V1.0 Operation Guide SETUP AND OPERATING PROCEDURES 1. BATTERY LOADING a. Open battery cover at the back of the monitor. b. Load two “AAA” size batteries. Please pay attention to polarity. c. Close the battery cover. When LCD shows battery symbol , replace all batteries with new ones. Rechargeable batteries are not suitable for this monitor. Remove the batteries if the monitor will not be used for a month or more to avoid relevant damage of battery leakage.
DN:ZRY721-SMSY01 V1.0 Picture 2-4 Operation Guide Picture 2-5 Picture 2-6 e. After adjusted clock and date,the LCD will blink “bt” see picture 2-7.Press M2 will alter START/STOP of bluetooth.If bluetooth is ON,a sign of will flash. Picture 2-7 f. During adjusting clock and date, the monitor will go back to Standby Mode automatically when no button will be pressed within 30 seconds. g.
DN:ZRY721-SMSY01 V1.0 Operation Guide 0:00 12:00 AM 12:00 12:00 PM 1:00 1:00 AM 13:00 1:00 PM 2:00 2:00 AM 14:00 2:00 PM 3:00 3:00 AM 15:00 3:00 PM 4:00 4:00 AM 16:00 4:00 PM 5:00 5:00 AM 17:00 5:00 PM 6:00 6:00 AM 18:00 6:00 PM 7:00 7:00 AM 19:00 7:00 PM 8:00 8:00 AM 20:00 8:00 PM 9:00 9:00 AM 21:00 9:00 PM 10:00 10:00 AM 22:00 10:00 PM 11:00 11:00 AM 23:00 11:00 PM 3.
DN:ZRY721-SMSY01 V1.0 Operation Guide to make sure that the appropriate cuff is used. • Measuring on same wrist each time. • Do not move your arm, body, or the monitor during measurement. • Stay still, calm for 5 minutes before blood pressure measurement. • Please keep the cuff clean. Clean the cuff by wet soft cloth and mild detergent if the cuff becomes dirty. Do not remove the cuff from the monitor.Clean the cuff after the usage of every 200 times is recommended. 5.
DN:ZRY721-SMSY01 V1.0 Operation Guide 7. DISPLAYING STORED RESULTS a. In StandBy Mode, press “MEM” button, the monitor will blink sign of current group.Press “MEM” button to switch group,press “START/STOP” to confirm current group.Then the amount of results in current user memory zone will be displayed. See picture 8. Then LCD will display the average value of all results in the current user memory zone.If the bluetooth function is on, the device will wait for pairing for 30s. See picture 8-1.
DN:ZRY721-SMSY01 V1.0 Note: Operation Guide When the monitor displaying the measurement, the classification color indicator can be shown different color according to the systolic pressure and diastolic pressure. Refer to the “ASSESSING HIGH BLOOD PRESSURE FOR ADULTS” section 8. DELETING MEASUREMENTS FROM THE MEMORY When any average value is displaying, keeping on pressing button “MEM” for three seconds, all results will be deleted after three “beep”.
DN:ZRY721-SMSY01 V1.0 Operation Guide The technical alarm condition (outside the rated range) is preset in the factory and cannot be adjusted or inactivated. This alarm condition is assigned as low priority according to IEC 60601-1-8. The technical alarm is non-latching and need no reset. The signal displayed on LCD will disappear automatically after about 8 seconds. 11.
DN:ZRY721-SMSY01 V1.0 Operation Guide MAINTENANCE 1. Do not drop this monitor or subject it to strong impact. 2. Avoid high temperature and solarization. Do not immerse the monitor in water as this will result in damage to the monitor. 3. If this monitor is stored near freezing, allow it to acclimate to room temperature before use. 4. 5. Do not attempt to disassemble this monitor. If you do not use the monitor for a long time, please remove the batteries. 6.
DN:ZRY721-SMSY01 V1.0 EC REP Operation Guide 47 Spenlow Bermondsey London SE16 4SJ Symbol forHouse “EUROPEAN REPRESENTATION” Symbol for “SERIAL NUMBER” Symbol for “KEEP DRY” WARRANTY INFORMATION Only charge the cost of components and transport. SERVICE CENTER ANDON HEALTH CO., LTD. No. 3 Jinping Street, YaAn Road, Nankai District, Tianjin 300190, China. Tel: 86-22-60526081 EC REP Lotus Global Co., Ltd.
DN:ZRY721-SMSY01 V1.0 Operation Guide For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacture’s declaration - electromagnetic emissions The ZRY721 is intended for use in the electromagnetic environment specified below. The customer or the user of the ZRY721 should assure that it is used in such an environment.
DN:ZRY721-SMSY01 V1.0 Operation Guide equation applicable to the frequency of the transmitter. Recommended separation Radiated RF 3 V/m 80 MHz to 2.5 IEC 61000-4-3 GHz 3 V/m distance: : 80 MHz to d = 1.2 P 800 MHz d = 2.3 P 800 MHz to 2.5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
DN:ZRY721-SMSY01 V1.0 Operation Guide b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. Table 4 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the ZRY721 The ZRY721 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.