User Manual

- 12 -
Inorganic metal oxide
Dilute acid less than 1mol/L
Organic lipid
Dilute detergent (weak alkaline)
Resin macromolecule
Dilute alcohol, acetone, ether
Proteinic haematocyte sediment
Acidic enzymatic solution (saccharated yeast tablets)
Paints
Dilute bleacher, peroxide
Notes:
1) The instruments will NOT give accurate and stable pH readings when measuring distilled or
deionized water. This because distilled and deionized water do not have enough ions present
for the electrode to function properly. To measure distilled or deionized water’s pH, users
need to use a specialized electrode. Contact us for more details.
2) The electrode housing is polycarbonate. When using cleaning solutions, take cautions on
carbon tetrachloride, trichlorethylene, tetrahydrofuran, acetone, etc., which will dissolve the
housing and invalidate the electrode.
5 mV MEASUREMENT
ORP measurement
Press key, and switch the meter to mV measurement mode. Connect ORP electrode (the 301Pt-S combination
ORP electrode is sold separately) and dip it in sample solution, stir the solution briefly and allow it to stay in the
solution until icon appears and get the reading.
ORP means Oxidation Reduction Potential. The unit is mV.
Notes on ORP measurement
5.2.1 ORP measurement does not require calibration. When the user is not sure about ORP electrode quality or
measuring value, use ORP standard solution to test mV value and see whether ORP electrode or meter works
properly. Table-6 is the data of standard ORP solution for 222 mV.
Table-6
°C
10
15
20
25
30
35
38
40
mV
242
235
227
222
215
209
205
201
5.2.2 Clean and activate ORP electrode
After the electrode has been used over a long period of time, the platinum surface will get polluted which causes
inaccurate measurement and slow response. Please refer to the following methods to clean and activate ORP