Technical data

AR2010 USB Stick Synthesizer USB Programming Manual
2. POWEr:SET [<numeric value>|MIN |MAX]
This command allows the user to set the output power of the AR2010 in 0.5 dB steps with a ±0.25 dBm ac-
curacy. The output power value is specified in dBm and rounded to the nearest 0.5 dBm. See your Prod-
uct’s Configuration Sheet for valid output power ranges. Selecting MAX or MIN configures the AR2010 to
output the maximum and minimum power levels, respectively, achievable at the current frequency. When
using Max or MIN no attempt is made to level the output power across frequency.
Syntax
POWE :SET Space numeric value|MIN|MAX
?
Example
POWE:SET -10 This command sets the output power of the device to -10 dBm
POWE:SET MAX This command sets the output power of the device to the maximum value achievable
at the current frequency
POWE:SET MIN This command sets the output power of the device to the minimum value achievable
at the current frequency
Default Condition
On power up, or when a SYST:PRES or *RST command is issued, the PLO output power defaults to the
power level stored in memory location 0, unless otherwise specified in the product’s Operation Manual.
Query
POWE:SET? This query returns the output power for the device in dBm.
Error Messages
If the power entered is out of the operational range of the device, an error message of 201, "Parameter
specified out of Device’s operating range" is placed in the device’s error queue.
Applied Radar, Inc. Revision 1.0.0 23