Technical data

AR2010 USB Stick Synthesizer USB Programming Manual
Indicates that a trigger command was received but ignored because the USB
Stick Synthesizer was not in the wait for trigger state.
-213 Trigger ignored
Indicates that a trigger command was received but ignored because the USB
Stick Synthesizer was not in the wait for trigger state.
-222 Data out of range
A numeric parameter value is outside the valid range for the command.
-224 Illegal parameter value
A discrete parameter was received which was not a valid choice for the com-
mand.
-230 Data corrupt or stale
This occurs when a measurement command is attempted and either a reset has
been received of the state of the USB Stick Synthesizer has changed such that
the measurement is no longer valid.
-241 Hardware missing
The USB Stick Synthesizer is unable to execute the command because the
hardware does not support that feature.
-310 System error
This error indicates a failure with the USB Stick Synthesizer.
-330 Self-test failed
The -330,"Self-test failed" error indicates a problem with the USB Stick Synthe-
sizer.
-350 Queue overflow
The error queue is full and another error has occurred which could not be
recorded.
-410 Query INTERRUPTED
A command was received which sends data to the output buffer, but the output
buffer contained data from a previous command. The output buffer is cleared
when power has been of or after a *RST or SYST:PRES command has been
issued.
-420 Query UNTERMINATED
The USB Stick Synthesizer was addressed to talk but a command has not been
received which sends data to the output buffer.
-430 Query DEADLOCKED
A command was received which generates too much data to fit in the output
buffer and the input buffer is also full. Command execution continues but data
is lost.
-440 Query UNTERMINATED after indefinite response
The *IDN? command must be the last query command within a command string.
+0 No error
Applied Radar, Inc. Revision 1.0.0 28