Technical data

USB Stick Synthesizer IEEE 488.2 Command Reference
6. *OPC
The *OPC (Operation Complete) command causes the AR2010 USB Stick Synthesizer to set the oper-
ation complete bit in the Standard Event Status Register when all pending device operations have been
completed.
Table 7.4: *OPC mapping
Bit Base 2 Meaning
0 1 Operation Complete
1 2 Not Used
2 4 Query Error
3 8 Device Dependent Error
4 16 Execution Error
5 32 Command Error
6 64 Not Used
7 128 Power On
Syntax
*OPC
?
Query
*OPC? This query places a 1 in the output queue when all device operations have been completed.
Applied Radar, Inc. Revision 1.0.0 48