User manual

AR2010 USB Stick Synthesizer User Manual
7. SYST:TEMPeratureTHRESHold <integer>
This command allows the user to set a maximum temperature threshold value for the USB Stick Syn-
thesizer in Celsius. If the on board sensor reads a temperature value above this threshold, the device
will power off the RF power supply. The digital power supply will remain enabled, allowing the user to
communicate with the device.
NOTICE
The SYST:TEMPTHRESH command is only available on the AR2010-5-10 v3.0.1 and higher RF Synthe-
sizer modules.
Syntax
SYST :TEMPTHRESH Space numeric value
?
Example
SYST:TEMPTHRESH 70 This command sets the temperature threshold to 70 degrees Celsius.
Query
SYST:TEMPTHRESH? This query returns the user deļ¬ned temperature threshold value.
Applied Radar, Inc. Revision 2.0.1 38