Operating instructions

AMC-DTR Sensor Transmitter User Manual
22
7.2 Configuration Using the Magnetic Wand
Passing the magnetic wand past the EDIT key, from either data display, displays SETUP PAGE
1 as shown in Figure 19. The UP / DOWN keys maneuver the pointer while EDIT enters sub-
levels of menu items. All SETUP menu items have at least one page of sub-menus. Items with
sub-menus are indicated by the > symbol (right hand pointing arrow) at the end of each line.
Edit menu items by pointing to them, press the EDIT key to display the cursor, press UP /
DOWN to change that character, press NEXT to move the cursor, then press EDIT again to load
the new item and remove the cursor. Press NEXT to reverse out of the sub-menu. To view
SETUP PAGE 2, press the DOWN key with the pointer aimed at the bottom item on PAGE 1.
Figure 19: Setup Menu Entry
7.3 System Configuration Menus
The System Config. group consists of two pages of menus as shown in Figure 20. Each item’s
description follows in this section.
S Y S T E M C ON F I G . P G 1
Me a s u r e me n t N a me -
16
E u n i t s X X X X X X X X X X -
10
Z e r o ( 0 % )
(Eunits)
S p a n ( 10 0 % )
(Eunits)
D e c i ma l P o i n t s 0
C a l S p a n V a l 5 0
Re a d o u t D e a d b a n d 0 %
T r a c k N e g a t i v e N O
L i n e a r i z a t i o n D a t a
B a c k u p C o n f i g .
*
R e s t o r e C o n f i g .
or
**
U p l o a d S e n s o r D a t a
UP/
DOWN
S Y S T E M C ON F I G. P G 2
*Simple Sensor Installed **Smart Sensor Installed
>
>
>
>
*“Restore” available only
without smart sensor installed.
**“Upload” available only
with smart sensor installed.
Figure 20: System Config. Menus
Measurement Name may be edited to contain virtually any 16-character ASCII field. It is
typically used to describe the monitored point by user tag # or other familiar terminology.
Eunits (engineering units) may have up to a 10 character ASCII field. Many common gases
have pre-configured Eunits based upon the sensor type and each may be edited in this menu as
described in Configuration Using the Magnetic Wand section 7.2.
Zero (0%) defines the reading to be displayed when 4mA (0%) is the AMC-DTR output.
Span (100%) defines the reading to be displayed when 20mA (100%) is the AMC-DTR output.
The highest reading allowed is 9999 includes negative polarity sign and one decimal point.
Polarity is only indicated for negative readings.
Decimal Points sets the resolution of the LCD readings and may be for 0, 1 or 2. Example:
ZERO readings for 0, 1 & 2 DP’s respectively are 0, 0.0 & 0.00.
Cal Span Value sets what upscale value must be applied when performing Span calibrations.