Owner's manual

1000V Digital/Analog Megohmmeter Models 1050 & 1060 (pdf) Rev. 05 01/05
Technical Assistance (800) 343-1391 www.aemc.com 2 of 9
FUNCTION MODEL MODEL
1050 1060
50V (0.002M to 200G) ✓✓
100V (0.004M to 400G) ✓✓
250V (0.01M to 1T) ✓✓
500V (0.02M to 2T) ✓✓
1000V (40k to 4T) ✓✓
Voltage Test/Safety Check (600VAC/DC) ✓✓
Guard Terminal ✓✓
Continuity (40) @200mA ✓✓
Resistance (400k) ✓✓
Lead Compensation ✓✓
Backlight ✓✓
Timer (1–59 minutes) ✓✓
Digital Dual Display ✓✓
Analog Bargraph ✓✓
Auto-Off ✓✓
Continuity Buzzer ✓✓
Continuity Buzzer Disable ✓✓
Voltage Detection (>25V) ✓✓
Safety Test Inhibitor ✓✓
Automatic Discharge ✓✓
Rechargable NiMH Battery
AC Supply 85-256V/50/60Hz
Battery Powered
Battery Capacity (%) Indication ✓✓
Auto-Off Disable ✓✓
Automatic DAR/PI ✓✓
Smooth Function ✓✓
Memory Storage (128kB)
Report Printout on Serial Printer
RS-232 Communication
Features
True Megohmmeter
®
Test voltage combinations of 50V, 100V, 250V,
500V and 1000V
Insulation measurements to 4000G (4T)
Direct measurement of DAR and PI values
Direct measurement of sample Capacitance
Display of test voltage and run time
Programmable test run times and PI times
Smooth and Alarm functions
Automatic test inhibition (if live sample >25V)
Automatic discharge after use with voltage display
Large dual display with time, voltage and measurement
Bright blue electroluminescent backlight
Battery powered
Auto power-down when not in use
Remote operation with optional test probe
Rugged dual wall field case with detachable lead/
accessory pouch
EN 61010-1, 600V Cat. III, EN 61557
CE Mark
Model 1060 includes these
additional features:
AC rechargeable NiMH batteries – run while recharging
RS-232 interface for direct printing of results (serial output)
128kB memory to store measurements in specific files
Remote operation of megohmmeter through PC
Includes DataView
®
software for data storage, real-time
display, analysis and report generation
Applications
Test insulation on cables, transformers, motors, insulators
and wiring installations
High resistance or absorption tests
Spot reading tests
Timed resistance measurements
Dielectric Absorption Ratio (DAR) and
Polarization Index (PI) tests
Low insulation test range for testing old or flooded
installations
Motor insulation resistance measurements
Continuity checks and low resistance measurements
Computer controlled production line testing
Predictive maintenance by storing results in PC for
trend analysis
Model 1060 checking insulation resistance between windings on a
three-phase motor.