Data Sheet

Released 08/2008 Rights reserv ed f or change in technical data! HYGROSENS INSTRUMENTS GmbH Postf ach 1054 D-79839 Löf f ingen Tel: +49 7654 808969-0 Fax: +49 7654 808969-9
CFriedrich, Dokument9
SHEATH RESISTANCE THERMOMETER
WITH FOUR WIRE CONNECTION M12
Characteristic features
Model PT100 or PT1000
Measuring range -70 ... +500 °C
High accuracy DIN class 1/3 B
Fast response behaviour
Four wire connection
Mechanically stable, flexible
Robust construction, stainless steel
Gas tight, hermetically sealed
Pressure and vacuum suitable
Simple mounting, 1/8“ threads
Industrial M12 plug, IP 67
Typical areas of application
Laboratory systems, science and research
Demanding industrial applications
Cryogenics and cooling systems
Plant manufacture
Features
Jacketed resistance thermometers combine the
accuracy of PT sensors with the advantages of
well proven jacketed thermocouples. The probe is
suitable for temperature range between -70... +500
°C and is the most optimum design with respect to
highest possible accuracy together with a robust
protected construction.
Mineral insulated jacketed cable is used as the
lead wire to sensor element. The lead wires are
embedded in a compact insulation of MgO and
enclosed in a metal jacket of high grade alloy non
rusting steel (material No. .: 1.4541 or INCONEL).
The compact insulation completely consolidates
the wire so that no damage can occur either due to
vibration or by bending pressure. The minimum
bending radius is around 15 mm.
The platinum sensor is placed at the tip of the
assembly in a protection sleeve which is hermeti-
cally tight sealed with the jacket lead wire. There-
fore, the model is very well suitable for critical
applications, for example, in the cryogenics. The
temperature sensitive length at the tip is 35 mm.
The protection sleeve is filled with MgO powder,
through which it has an extremely fast conversa-
tion behaviour with optimum protection to the
sensor, for example in case of vibration.
The mounting in the installation is either done
through the 1/8 “ threads or 6mm clamping set.
The connection to M12 plug connector is through a
four wire system. The plug connector has a protec-
tion type IP67. The scope of supply is without
connection cable; if required, please separately
include with the order!
Technical data
Transducer module CON-MANTE
Measuring range -70 ... +500 °C
Sensor element Platinum thin film sensor
Accuracy DIN Class 1/3 B
from 0 ... +100 °C
Response time t90 8 seconds (in oil)
Resistance
characteristics
PT100 or PT 1000 as per
DIN EN 60751
Connection Four wire connection
Probe tube Length 100 mm, hermetically
sealed jacket tube, stainless
steel grade 1.4541
Minimum bending
radius
15 mm
Gland Stainless steel 1.4305
Dimensions SW 14 x 148 mm
Internal threads 1/8“ with built in gasket
Gasket DIN G1/8“ x 10
Connection plug M12 connection plug,
4-pin (IP67)
Temperature at plug -20 ... +80 °C
Further models with modified jacket tube length or
other sensor elements are available on enquiry.
Rights reserved for change in technical data!
For further information, please visit our
website:
www.hygrosens.com
4 0161 38 48 9 1 1 8
Ordering No. 502144
CON-MANTE-PT100

Summary of content (2 pages)