User Manual

87
8 Auto Measurement
The oscilloscope measures 36 waveform parameters and can present statistics on up to five of
these parameters. These measurements are voltage, time and delay parameters.
To set up the measurement function:
1. Press Measure to enter the MEASURE menu.
2. If desired, press the Clear softkey to remove any existing measurement parameters.
3. Press the Source softkey to choose the waveform to measure. CH1, CH2, MATH, REF A, or
REF B.
4. Press the Type softkey and turn the Universal Knob to select the desired measurement
parameter(s). The trigger Level Knob can also be used to move the selection.
5. Press the Universal Knob to add the selected parameter to the list of five measurements to
be displayed.
6. Press the Type softkey to exit the parameter selection window.
The measurements will be arranged from left to right according to the selection order, first in,
first out.
Figure 85 - Selecting the measurement parameter
Note: If the parameter is unable to be measured, it will display as ****.