User Manual

102
11 Utility
The System Function menu allows the user to view system-related information and access
functionality such as system status, performing a self-calibration, sound, language, pass/fail
testing, configuring the I/O interfaces, updating the firmware and configuration, setting the
screen saver, and configuring optional features.
11.1 View the System Status
1. Press the Utility button to enter the UTILITY menu.
2. Press the System Status softkey to view the system status of the oscilloscope. The system
status includes the information:
Startup Times: Number of times the oscilloscope has been powered on.
Software Version: Lists the current software version of the oscilloscope.
FPGA Version: Lists the current FPGA version of the oscilloscope.
Hardware Version: Lists the current hardware version of the oscilloscope.
Product Type: Gives the model number of the oscilloscope.
Serial No.: Lists the serial number of the oscilloscope.
Scope ID: Displays the oscilloscope identifier.
3. Press the Single button on the front panel to exit this screen.
11.2 Self Calibration
The self-calibration program lets the oscilloscope provide its most precise measurement values.
You can perform self-calibration at any time. A self-calibration is recommended when ambient
temperature has changed 5 °C or more. Make sure that the oscilloscope has been turned on for
more than 30 minutes before performing the self-calibration. It will take approximately 5 minutes
for the self-calibration to complete.
1. Remove all connections from all input channels.
2. Press the Utility button.