User Manual

123
Red dots in the decode line indicate that more data can be displayed. Scroll or expand the
horizontal scale to view the data.
Interpreting I2C Lister Data
In addition to the standard Time column, the I2C Lister contains these columns:
NO: frame number from left to right.
TIME: time for each frame.
ADDRESS: colored blue for writes, yellow for reads.
R/W: yellow R for read, dark green W for write and black X for missing.
DATA: data bytes.
15.2 SPI Serial Decode
The following order of operations should be followed to ensure that the trigger and decode
functions are properly set:
Setup for SPI Signals
SPI Decode
Setup for SPI Signals
Serial Peripheral Interface (SPI) signals setup consists of connecting the oscilloscope to a clock,
MOSI data, MISO data, and framing signal, then setting the threshold voltage level for each input
channel, and finally specifying any other signal parameters.
To set up the oscilloscope to capture SPI signals:
1. Press the Decode button to enter the DECODE menu.
2. Press the Decode softkey to select the desired slot (Decode 1 or Decode 2).
3. Press the Protocol softkey and select SPI with the Universal Knob and press the knob to
confirm.
4. Press the Signal softkey to enter the SPI SIGNAL menu.
Figure 109 - SPI SIGNAL menu
5. Press the CLK softkey to enter CLK (clock) menu.
Figure 110 - SPI CLK menu
In the SPI CLK menu:
a. Press the CLK softkey.