User Manual

147
10. Press the Setting softkey to enter the SETTING menu.
11. Press the Output Load softkey to select high impedance (High-Z) or 50 Ω.
17.3 Output Impedance
1. Press the Wave Gen button to enter the WAVEFORM menu.
2. Press the Setting softkey to enter the SETTING menu.
3. Press the Output Load softkey to select high impedance (High-Z) or 50 Ω. The impedance
selection will be shown on the top line of the display next to Awg.
17.4 Set Default Values
A softkey in the WAVEFORM menu lets you restore the default wave type, frequency, amplitude,
and DC offset:
1. Press the Wave Gen button to enter the WAVEFORM menu.
2. Press the Setting softkey to enter the SETTING menu.
3. Press the Default softkey to restore the default values a sine waveform of 1 kHz frequency, 4
Vpp, and 0 V DC offset.
17.5 AWG Self Cal
If the ambient operating temperature changes more than 5 °C and the instrument runs more
than 30 minutes, you should run the AWG self calibration routine to reduce thermal offset
effects.
1. Press the Wave Gen button to enter the WAVEFORM menu.
2. Press the Setting softkey to enter the SETTING menu.
3. Press the AWG Self Cal softkey to run the AWG self-calibration program.
Note: The output impedance of the generator must match that of the oscilloscope or the amplitude of the
signal will be incorrect.