User Manual

7
Icon
Function
Description
Trigger Type
Displays the currently selected trigger type and trigger
condition setting. Different labels are displayed when
different trigger types are selected.
Trigger Source
Displays the trigger source currently selected. Different
labels are displayed when different trigger source are
selected and the color of the trigger parameter area will
change accordingly.
Trigger Slope
Displays the current trigger slope.
Trigger Coupling
Displays the coupling mode (DC/AC/LF Reject/HF Reject)
of the current trigger source.
Trigger Level
Displays the trigger voltage or current level of the
current waveform. Press the Universal Knob to set the
level to 50% of the waveform's amplitude.
Table 1 - Trigger settings
8. Channel setting
The channel settings are displayed when the represented channel is enabled. If no channel is
enabled there will be no channel setting display.
Figure 6 - Channel settings display
Icon
Function
Description
Channel Number
Represents the channel number
Input impedance
Displays the currently selected input impedance of
the channel (1 MΩ or 50 Ω).
Channel coupling
Displays the coupling mode of the current channel.
The modes are DC, AC, and GND.