Specifications

Halogen Gas Monitor – MultiZone Remote Display
Instruction 3015-5157 21
Navigating to the 2
nd
RDM Setup Screen
On RDM Setup Screen #1, select MORE to access RDM Setup Screen #2. Select BACK to return to RDM
Setup Screen #1.
RDM Setup Screen #1 RDM Setup Screen #2
Setting Relay Parameters
This setting determines the function of relays 1 and 2 mounted inside the RDM unit.
1. Press the ENTER key while the relay parameter you want to adjust is highlighted.
2. Use the UP/DOWN cursor keys to select Unused, Monitor On, Evacuate, Spill, Leak, Fault, or
Alarm.
3. Press ENTER to accept the new entry or ESC to cancel.
HGM Node Address
This value is the node address from 1 to 15 assigned to each HGM-MZ which is on the network. These
values must match the settings on each HGM-MZ.
1. Press the ENTER key to adjust the address for each HGM-MZ.
2. Use the UP/DOWN cursor keys to modify the setting.
3. Press ENTER to accept the new entry or ESC to revert to the previous setting.