Specifications

Halogen Gas Monitor – MultiZone Remote Display
Instruction 3015-5157 24
Re-Zero Mode
This parameter defines the frequency at which the instrument re-zeros the optical sensor.
1. Press the ENTER key to modify the setting.
2. Use the UP/DOWN cursor keys to toggle between settings.
AUTO Sets the instrument to re-zero every 10 minutes.
ZONE CHANGE Sets the instrument to re-zero at each zone change. This is the most
accurate setting but increases the time interval between measurement cycles.
3. Press ENTER to accept the new entry or ESC to revert to the previous setting.
Navigating to the 3
rd
HGM Setup Screen
From HGM Setup Screen #2, select the MORE option to go to HGM Setup Screen #3. Select the BACK
option to return to HGM Setup Screen #2. To return to the System Screen, press ESC.
HGM Setup Screen #2 HGM Setup Screen #3
Service Timeout
Sets the length of time the Service Mode is in effect. Stops the monitor for up to 240 minutes (4 hrs) to allow
for servicing the unit. Refer to Service Mode on Page 46.
1. Press the ENTER key to modify the setting.
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to
modify the individual numbers.
3. Press ENTER to accept the new entry or ESC to revert to the previous setting.