Specifications

Halogen Gas Monitor – MultiZone Remote Display
Instruction 3015-5157 30
Alarm Conditions
When an alarm condition is detected anywhere on the network the red ALARM LED will glow. Additionally, an
external alarm device may activate and an audible alarm may sound if those features have been enabled
(Pages 21 & 25).
An inverse flashing box indicates an alarm condition in the affected zone.
Zone Screen (Alarm Mode)
If the current Zone Screen appears normal, scroll through the displays using the NEXT/BACK keys or the
UP/DOWN cursor keys until you locate the affected zone. To further investigate the status of an affected
zone, select the zone and press the ENTER key. That Zone’s Setup Screen #1 will be displayed. You may
also select ALARMS to access the Alarm Summary Screen (Page 34). Refer to Working with Alarms on
Page 34 to learn more about responding to alarms.
Fault Conditions
If a system malfunction occurs, the yellow FAULT LED will glow. Additionally, an external alarm device may
activate and an audible alarm may sound if those features have been enabled (Pages 21 & 25).
From the top level Zone Screen (Page 30), make sure no zones are selected and pres ENTER to go to the
System Screen (Page 19) and then press the FAULTS key to go to the Fault Screen (Page 38). Refer to
System Faults on Page 39 to learn more about responding to system faults.
Fault Screen
Zone in Alarm
(TBD)