Bair Hugger Model 775 Temperature Management Unit Service Manual
English 202526A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
Overview
When you connect the Bair Hugger Model 775 Temperature Management Unit to a grounded power source,
the unit automatically performs the power-on-reset operation described below� Following the power-on-
reset operation, the unit will enter the Standby mode� You can select any operating mode by pressing the
appropriate button on the unit’s control panel� See Figure 1�
Figure 1. Control panel
Model 775
43
˚C
STANDBY
TEMP IN
RANGE
FAU LT
OVER
TEMP
AMBIENT
H
I
G
H
L
O
W
M
E
D
High Fan
Setting
Low Fan
Setting
Temperature Modes
Unit Power-On-Reset
e Bair Hugger Model 775 temperature management unit automatically performs the following power-on-
reset sequence aer you connect the unit to a properly grounded power source:
• Performs all self-test functions�
• Illuminates indicator lights and all pixels in the alphanumeric display momentarily
• Displays the text “MD 700’s” and the soware revision level in the alphanumeric display
• Sounds the alarm (three low-level clicks)�
• Enters the Standby mode with the high fan ( ) setting selected by default
note: e low fan setting indicator light will not illuminate during power up
If the Bair Hugger Model 775 Temperature Management Unit loses power for less than 1 second, the units
soware will return the unit to the operating mode you selected prior to the power loss� However, if the
temperature management unit loses power for longer than 1 second, the unit’s soware will reset when you
restore power� e unit will then enter the Standby mode with the high fan setting selected by default�