Service manual
9
Bair Hugger Model 775 Temperature Management Unit Service Manual
English 202526A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
Fault Conditions
e Model 775 Temperature Management Unit’s soware recognizes several nonhazardous fault conditions
including:
• Sensor failures
• Heater failures
• Computer failure
• Soware failure
• Keypad failure
• Power supply failure
• Blower motor failure
• Forced over-temperature test failure
When a fault condition exists, the following events occur:
• Fault indicator light ashes�
• Audible alarm sounds�
• Blower heater turns o�
• Fault code appears in the alphanumeric display�
• Up to 5 fault codes are stored in memory�
When the audible alarm sounds, press any button on the keypad to temporarily silence it� e temperature
management unit must be unplugged before it can be restarted� See "Viewing the Fault Code Log" on
page16 and the "Fault Code Table" on page17 for more information about specic fault codes�
Over-Temperature Condition
When an over temperature condition exists, the following events occur:
• Over-temp indicator light ashes�
• Audible alarm sounds�
• Blower and heater turn o�
• Fault code appears in the alphanumeric display (see page 17)�
When the audible alarm sounds, press any button on the keypad to temporarily silence it� e temperature
management unit must be unplugged before it can be restarted�