Method
N: If the Bair Hugger Model 775 Temperature Management Unit has been stored at a temperature lower
than 20°C, allow it to reach ambient temperature before beginning the calibration procedure�
1� Attach the Model 22110 temperature test device to the end of the temperature management unit hose�
2� Connect the temperature management unit to a grounded power source� e unit performs a self-test
and automatically proceeds to Standby mode�
3� Press the Ambient button to place the unit in Ambient mode� Allow the unit to run in Ambient mode
for at least ten minutes to equalize the temperature of the unit’s internal components�
Note: Verify unit is set to high fan setting by ensuring the indicator light next to the upper fan is
illuminated�
4� Press the Alternative Mode button (see "Figure 2� Location of Alternative Mode button" on
page11) and the Ambient button simultaneously for 3 seconds� e temperature management unit
is now in Calibration mode� In this mode, the temperature management unit performs the following
tasks:
• 32°C and 38°C indicator lights ash�
• Blower remains ON
• Heater elements remain OFF
• e text “TMP” and the end of hose temperature in °C (0�1°C resolution) appear in the
alphanumeric display
• e text “OS” and the current calibration oset value (0�1°C resolution) appear in the
alphanumeric display
5� Compare the temperature shown on the alphanumeric display and the temperature of the Model
22110, temperature test device
• If the temperatures match, press the Alternative Mode button to return the temperature
management unit to Standby mode
• If the temperatures do not match, adjust the temperature management unit’s displayed
temperature:
a� Press the 32°C button to decrease the temperature or the 38°C button to increase the
temperature by 0�1°C� You can adjust the temperature of the temperature management unit a
maximum of ±3°C�
b� Continue to press the 32°C or 38°C button(s) until the temperature in the alphanumeric display
matches the temperature on the calibration device as close as possible�
c� Press the Alternative Mode button to store the calibration oset value and return the
temperature management unit to Standby mode�
N: If it is necessary to enter an oset value greater than ±3°C, there may be a defect in the hose end
sensor� Remove unit from service until a new hose assembly can be substituted�